5IRE
The cryo-EM structure of Zika Virus
ELECTRON MICROSCOPY
Sample |
---|
Zika virus |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | GATAN CRYOPLUNGE 3 |
Cryogen Name | ETHANE |
Sample Vitrification Details | Blot for 5 seconds before plunging into liquid ethane (GATAN CRYOPLUNGE 3). |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 11842 |
Reported Resolution (Å) | 3.8 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | Soft mark was applied to the even/odd map before FSC calculation. |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | I |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | AB INITIO MODEL | ||||
Refinement Target | T = Tmap + weight * TrestraintsT = Tmap + weight * Trestraints | ||||
Overall B Value | 150 | ||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 23 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 1000 |
Maximum Defocus (nm) | 2500 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 14000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | Appion | 3.2 |
IMAGE ACQUISITION | Leginon | |
CTF CORRECTION | CTFFIND | 3 |
MODEL REFINEMENT | PHENIX | 1.10.1-2155 |
INITIAL EULER ASSIGNMENT | jspr | 2015 |
FINAL EULER ASSIGNMENT | jspr | 2015 |
CLASSIFICATION | jspr | 2015 |
RECONSTRUCTION | jspr | 2015 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING ONLY | Phase flipping was performed during 3D reconstruction. | 64518 | Particles were picked from 2-binned images using the template-picking method from Appion. |